Search

📱Telephone:400-686-3443

中文  |  EN

D-虫荧光素游离酸

Luciferin is a common bioluminescent reporter used for in vivo imaging of the expression of luciferase. This water soluble substrate for the firefly luciferase enzyme utilizes ATP and Mg2+ as cofactors to emit a characteristic yellow-green emission in the presence of oxygen, which shifts to red light in vivo at 37°C. Through the utilization of ATP, the reaction can be further used to indicate the presence of energy or life in order to function as a life-death stain.
零售价
0.0
市场价
0.0
浏览量:
1000
产品编号
12501-25mg/400元;100mg/1000元
数量
-
+
库存:
0
暂时无货
1
产品描述
[[[[[[[[[[[[[[[[[[[[[[[[产品参数, 参数]]]]]]]]]]]]]]]]]]]]]]]]

D-虫荧光素游离酸 [D-Luciferin, free acid]

 

Chemical Name:D-Luciferin, free acid虫COOH.png

MW:C11H8N2O3S2=280.32; CAS#2591-17-5

Luciferin is a common bioluminescent reporter used for in vivo imaging of the expression of luciferase. This water soluble substrate for the firefly luciferase enzyme utilizes ATP and Mg2+ as cofactors to emit a characteristic yellow-green emission in the presence of oxygen, which shifts to red light in vivo at 37°C. Through the utilization of ATP, the reaction can be further used to indicate the presence of energy or life in order to function as a life-death stain.

Luciferin is a common reagent used throughout the biotechnology field and specifically for in vivo imaging. Luciferase labeled tumor cells, stem cells or infectious diseases are often inoculated into research animals such as rats or mice for investigation. The injection of luciferin allows for the real-time, noninvasive monitoring of disease progression and/or drug efficacy in these model systems through Bioluminescence Imaging (BLI).

Luciferin is also commonly used for in vitro research, including luciferase and ATP assays, gene reporter assays, high throughput sequencing and various contamination assays.

 

 

Preparation of Luciferin for In Vitro and In Vivo Bioluminescent Assays

 

Preparation of Luciferin for In Vitro Bioluminescent Assays

Materials: D-Luciferin Firefly, potassium salt, 1.0 g /vial; Sterile water; Complete media

 

Cells should be seeded in a plate overnight or several hours prior to assay to allow the cells to attach to the bottom. Suspension cell lines can be seeded in the working solution in the plate for direct incubation and imaging.

Doubling time should be considered for cell counting if doubling time is relatively short.

 

Procedure

 

Prepare a 200X Luciferin stock solution (30 mg/ml) in sterile water. Note: One can either reconstitute the entire 1.0 g of D-Luciferin in 33.3 mL of sterile water to make the 30 mg/mL (200x) stock solution, or reconstitute the quantity of D-Luciferin necessary for an individual experiment.

 

Mix gently by inversion until Luciferin is completely dissolved. *

 

Prepare a 150 µg/ml working solution of D-Luciferin in pre-warmed tissue culture medium. Quick thaw 200X stock solution of Luciferin and dilute 1:200 in complete media (150 µg/ml final concentration).

 

Aspirate media from cultured cells.

 

Add 1x Luciferin solution to cells just prior to imaging. Note: Incubating the cells for a short time at 37 °C before imaging can increase the signal. Incubation time is dependent on the specific cell type. Generally, a 10 min incubation is sufficient; test and adjust as needed.

 

Check the in vitro bioluminescence using the IVIS imaging system every 10 min, up to 40 min, to determine the kinetic curve and find the peak imaging time point for each cell type.

 

Preparation of Luciferin for In Vivo Bioluminescent Assays

 

Materials

 

D-Luciferin, Firefly, potassium salt, 1 g/vial 

DPBS, w/o Mg2+ and Ca2+

Syringe filter, 0.2 um

 

Procedure

 

Prepare a fresh stock solution of Luciferin at 15mg/ml in DPBS. *

 

Filter sterilize through a 0.2 um filter.

 

Determine injection amount at 10 µL/g of body weight. Each mouse should receive 150 mg Luciferin/kg body weight. (e.g. For a 10 g mouse, inject 100 µL to deliver 1.5 mg of Luciferin.)

 

Inject the Luciferin intra-peritoneally (i.p.) 10-15 minutes before in vivo imaging, or as determined by kinetic curve. **

 

* Immediate use of working solution after dilution is strongly recommended. If necessary, dissolved luciferin may be stored at 4°C for up to 3 weeks; however prolonged storage at either 4°C or -20°C may result in degradation of signal.

 

** A Luciferin kinetic curve should be performed for each new animal model to determine peak signal time. Please see our ‘Determining the Luciferin Kinetic Curve for Your Model’ instruction sheet available for download on our website.

 

NOTE: Luciferin is a light sensitive reagent, and should be kept out of direct light as much as possible.

We recommend that the luciferin be protected from light (e.g. covered with a light blocking material such as tin foil) during the entire assay, from the stock solution preparation until completion of the procedure.

 

Safety warnings and precautions 

 NOTE: The qualified persons should always wear lab coats, gloves and goggles when working with our products although they are low-risk chemicals for R&D only.

 

特别提醒:

1) 本公司所有产品仅限于专业人员用于生命科学研究,不得用于临床诊断或治疗,不得用于食品或药品,不得存放于普通住宅。

2) 本公司所有产品必须由合格专业技术人员操作同时佩戴口罩/手套/实验服并遵守生物实验室安全操作规程!

 

Description

Name ​D-Luciferin, free acid
CAT# 12501 CAS# 2591-17-5
Storage# −20°C Sealed & desiccated & Minimized light exposure Shelf Life# 12个月
Ex(nm)# 328 Em(nm)# 533
MW# 280.32 Solvent# DMSO
扫二维码用手机看
未找到相应参数组,请于后台属性模板中添加

相关企业下载

There is currently no content to display
Please add data record on website background.
底部logo

The ultimate scientific research that strives for excellence is manifested in the details, forging the beauty of product quality.
Day after day, we persist in the pursuit of innovation. Through dedication, we experience and achieve a leading position in technology.

imgboxbg

Room 208, 2nd Floor, Building 2, No. 4 Courtyard, Jinhang West Road, Shunyi District, Beijing

Customer Service hotline:400-686-3443